datasheetbank_Logo
データシート検索エンジンとフリーデータシート

SDX05G2-A データシートの表示(PDF) - Sensortechnics GmbH

部品番号
コンポーネント説明
一致するリスト
SDX05G2-A Datasheet PDF : 4 Pages
1 2 3 4
SDX...A Series
Precision compensated pressure sensors in DIP package
PRESSURE SENSOR CHARACTERISTICS (all devices)
Maximum ratings
Supply voltage VS
Maxium pressure on any port
Lead temperature (soldering 4 sec.)
+20 VDC
150 psig
250°C
STANDARD PRESSURE RANGES
Environmental specifications
Temperature ranges
Compensated
Operating
Storage
Humidity limits
Part number
SDX01...
SDX05...
SDX15...
SDX30...
SDX100...
Operating
pressure
0 - 1 psi
0 - 5 psi
0 - 15 psi
0 - 30 psi
0 - 100 psi
Proof11
pressure
20 psi
20 psi
30 psi
60 psi
150 psi
Min.
17.82 mV
59.40 mV
89.10 mV
89.10 mV
99.00 mV
Full scale span1,2
Typ.
18.0 mV
60.0 mV
90.0 mV
90.0 mV
100.0 mV
0 to 50°C
-40 to +85°C
-55 to +125°C
0 to 100% RH
Max.
18.18 mV
60.60 mV
90.90 mV
90.90 mV
101.00 mV
COMMON PERFORMANCE CHARACTERISTICS1
Characteristics
Zero pressure offset
Combined linearity and hysteresis3
Temperature effects (0 to 50°C)4
Repeatability5
Input impedance6
Output impedance7
Common mode voltage8
Response time9
Long term stability of offset and span10
Offset
Span
Min.
-0.300
1.5
Typ.
0
±0.1
±0.2
±0.4
±0.2
4.0
4.0
3.0
100
±0.1
Max.
+0.300
±0.25
±0.5
±1.0
±0.5
5.0
Unit
mV
%FSO
mV
%FSO
k
VDC
µsec
mV
Specification notes:
1. Reference conditions (unless otherwise noted): supply voltage, VS = 12VDC; TA = 25°C; common mode line pressure = 0 psig;
pressure applied to port 2. For absolute devices only, pressure is applied to port 1 and the output polarity is reversed.
2. Span is the algebraic difference between the output voltage at full scale pressure and the output at zero pressure. Span is
ratiometric to the supply voltage.
3. Hysteresis is the maximum output difference at any point within the operating pressure range for increasing and decreasing
pressure. Maximum linearity & hysteresis for the SDX05 is 0.5 %FSO.
4. Maximum error band of the offset voltage and the error band of the span, relative to the 25°C reading.
5. Maximum difference in output at any pressure within the operating pressure range and temperature within 0 to +50°C after:
a) 100 temperature cycles, 0 to +50°C.
b) 1.0 million pressure cycles, 0 psi to full scale span.
6. Input impedance is the impedance between VS and ground.
7. Output impedance is the impedance between + and - outputs.
8. This is the common mode voltage of the output arms for VS = 12 VDC.
9. Response time for a 0 psi to full scale span pressure step change, 10 to 90 % rise time.
10. Long term stability over a one year period.
11. If the maximum pressure is exceeded, even momentarily, the package may leak or burst, or the pressure sensing die may fracture.
The proof pressure for the forward gage of all devices in the D4-package is the specified value or 100 psi, whatever is less.
June 2008 / 031
2/4
www.sensortechnics.com

Share Link: 