MAX31865
RTD-to-Digital Converter
3-Wire (D4)
Write 1 to this bit when using a 3-wire RTD connection.
In this mode the voltage between FORCE+ and RTDIN+
is subtracted from (RTDIN+ - RTDIN-) to compensate
for the IR errors caused by using a single wire for the
FORCE- and RTDIN- connections. When using 2-wire or
4-wire connections, write 0 to this bit.
Fault Detection Cycle (D3:D2)
The master initiated fault detection cycle has two modes
of operation, manual and automatic mode timing. If the
external RTD interface circuitry includes an input filter with
a time constant greater than 100Fs, the fault detection
cycle timing should be controlled in the manual mode
operation. The fault detection cycle checks for three
faults by making the following voltage comparisons and
setting the associated bits in the Fault Status Register:
1) Is the voltage at REFIN- greater than 85% x VBIAS?
(Fault Status Register bit D5)
2) Is the voltage at REFIN- less than 85% x VBIAS when
FORCE- input switch is open? (Fault Status Register
bit D4)
3) Is the voltage at RTDIN- less than 85% x VBIAS when
FORCE- input switch is open? (Fault Status Register
bit D3)
Note: All voltages are referenced to GND1.
The Applications Information provides tables for decoding
possible causes of set fault status bits.
To enter the automatic fault detection cycle, write
100X010Xb to the Configuration register. The ADC
is now in “Normally Off” mode. The automatic fault
detection cycle inserts 100Fs delays before checking for
faults, thereby allowing the external input filter to settle.
The Fault Detect Cycle bits (D[3:2]) self-clear to 00b
upon completion.
To enter the manual fault detection cycle, first ensure that
VBIAS has been on for at least 5 time constants. Next,
write 100X100Xb to the Configuration register. The ADC
is now in “Normally Off” mode. The MAX31865 checks
for faults while the FORCE- input switch is closed, and
when the check completes, the FORCE-input switch
opens. The Fault Detect Cycle bits (D[3:2]), remain set to
10b. Again, wait at least 5 time constants, and then write
100X110Xb to the Configuration register. The MAX31865
now checks for faults while the FORCE- inputs switch
is open; when the check completes, the FORCE- input
switch closes and the Fault Detect Cycle bits (D[3:2])
self-clear to 00b. Note that if 1 is written to D5 (1-Shot)
and D2 or D3 in a single write, both commands are
ignored. If 100X110Xb is set without a prior initiation of
the first manual step (setting 100X100Xb), the automatic
fault detection mode is run instead.
Fault Status Clear (D1)
Write a 1 to this bit while writing 0 to bits D5, D3, and D2
to return all fault status bits (D[7:2]) in the Fault Status
Register to 0. Note that bit D2 in the Fault Register, and
subsequently bit D0 in the RTD LSB register may be set
again immediately after resetting if an over/undervoltage
fault persists. The fault status clear bit D1, self-clears to
0.
50/60Hz (D0)
This bit selects the notch frequencies for the noise
rejection filter. Write 0 to this bit to reject 60Hz and
its harmonics; write 1 to this bit to reject 50Hz and its
harmonics. Note: Do not change the notch frequency
while in auto conversion mode.
Table 3. Fault-Detection Cycle Control Bits
D3 D2
00
01
10
CONFIGURATION REGISTER
WRITE (BINARY)
XXXX00XXb
100X010Xb
100X100Xb
11
X = Don’t care
100X110Xb
WRITE ACTION
READ MEANING
No action
Fault detection with automatic delay
Run fault detection with manual delay
(cycle 1)
Finish fault detection with manual delay
(cycle 2)
Fault detection finished
Automatic fault detection still running
Manual cycle 1 still running; waiting for
user to write 11
Manual cycle 2 still running
Maxim Integrated
  13