SPECIFICATIONS
q Main unit
Type No.
Photodetector
No. of photosensitive device channels
Channel size
Cooling temepareture
Read-out noise
Dark current
A/D resolution
Spectrograph F number
Spectrograph type
Simultaneous measurement
wavelength range
C5965-31
MOS linear image sensor
1024 ch
25 µm(H) × 2.5 mm (V)
non-cooling
10 000 electrons
12,500 electrons/scan ( at 25˚C; 20ms )
C5966-3x
CCD linear image sensor
1024 ch
24µm (H) × 3.07 mm(V)
0˚C
60 electrons
512 electrons/scan( at 0˚C; 20ms )
3
Concave spherical grating type
300 nm to 800nm
x=1 300 nm to 800 nm
x=2 200 nm to 400 nm
x=3 600 nm to 1000nm
C7473-36
BT- CCD linear image sensor
1024 ch
24 µm(H) × 2.928 mm(V)
-15˚C
10 electrons
75 electrons/scan ( at -15˚C; 20ms )
16bit
200 nm to 950 nm
Wavelength resolution *
< 3 nm(FWHM)
x=1 < 3 nm(FWHM)
x=2 < 1.5 nm(FWHM)
x=3 < 2.5 nm(FWHM)
< 2 nm(FWHM)
Effective Light-receiving area of optical fiber
1mm
Optical fiber length
1.5m
Exposure time
20ms to 32s
External trigger input
TTL level / High impedance
Interface
SCSI
Line voltage
AC100V to 240V ±10%, 50, 60Hz
*Tested by the bright-line spectrum of Hg-Ar lamp (at 312.57nm, 435.84nm, 546.07nm, 696.54nm, 1013.98nm)
C8147-34
C8147-38
InGaAs linear image sensor
256 ch
50 µm(H) × 250 µm (V)
0˚C
-10˚C
12,500 electrons
20,000 electrons/scan ( at 0˚C; 5ms ) 2.5 × 107 electrons/scan ( at -10˚C; 5ms )
4
Czerny-Turner type
900 nm to 1650 nm
1600 nm to 2350 nm
< 9 nm(FWHM)
< 9 nm(FWHM)
5 ms to 32 s
5 ms to 50 ms (typ.)
q Basic software
Measurement
functions
Temporal resolution
measurement functions
Data acquisition
condition setting
Calibration and
correction
Display functions
Wavelength axis display
Brightness axis display
Cursor analysis
functions
Other analytical
functions
• Spectral measurement
• Reflection spectra measurement
• Absorption spectra measurement
• Color measurement
• Temporal fluctuation of spectra over time
• Temporal fluctuation of reflection factor and
transmittance over time
• Exposure time
• Memory integration count times
• Temporal fluctuation measurement
• Wavelength axis
• Sensitivity uniformity
• Dark current
• Spectrum (non-limited accumulation)
• Temporal fluctuation of waveform over time
(non-limited accumulation)
• Chromaticity diagram
• Wavelength (nm) ,Wavenumber (cm-1),
Energy (eV)
• Linear, logarithm
• Wavelength (Wavenumber etc.) vs, intensity
• Peak detection
• FWHM between two cursors
• Integrated intensity
• Smoothing
• Differential waveform
• Color measurement
DIMENSIONAL OUTLINES (Unit :mm)
q C5965. C5966 (approx. 4.5kg)
POWER
PHOTONIC MULTI-CHANNEL ANALYZER
SIGNAL INPUT
PMA-11
170±1
232±1
3±0.5 40±1
q C7473 (approx. 5kg)
POWER
PHOTONIC MULTI-CHANNEL ANALYZER
SIGNAL INPUT
C7473
260±1
333±1
200±1
262±1
3±0.5 25±1
q C8147 (approx. 7.5kg)
PHOTONIC MULTI-CHANNEL ANALYZER
PMA-11
SIGNAL INPUT
POWER
325±1
383±1
200±1
262±1
3±0.5 25±1
q Fiber input optics (approx.100g)
PVC sheathed SUS
flexible tube ( 7)
C1
375±1
433±1
3
Bundle diameter 1.0
80
50
15 strands
1500
ISO 9001
Certificate: 09 105 79045
5 Product and software package names noted in this documentation are trademarks or registered trademarks of their respective manufacturers.
q Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office..
q Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications and external appearance are subject to change without notice.
© 2001Hamamatsu Photonics K.K.
Homepage Address http://www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Systems Division
812 Joko-cho, Hamamatsu City, 431-3196, Japan, Telephone: (81)53-431-0124, Fax: (81)53-435-1574, E-mail:export@sys.hpk.co.jp
U.S.A. and Canada: Hamamatsu Photonic Systems: 360 Foothill Road, Bridgewater, N.J. 08807-0910, U.S.A., Telephone: (1)908-231-1116, Fax: (1)908-231-0852, E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658, E-mail: info@hamamatsu.de
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10, E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10Tewin Road,Welwyn Garden City, Hertfordshire, AL7 1BW, U.K., Telephone: (44) 1707-294888, Fax: (44) 1707-325777, E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 Solna, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01, E-mail: info@hamamatsu.se
Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Mois, 1/E 20020 Arese (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741, E-mail: info@hamamatsu.it
Cat. No. SSCS1040E07
JUL /2002 HPK
Created in Japan (PDF)